Home / Environmental Test Equipment / Temperature & Humidity / Benchtop Humidity Chamber / SM-1.0-8200 Benchtop Environmental Chamber
SM-1.0-8200 Benchtop Environmental Chamber
Workspace Dimensions—WxDxH: |
14 x 8 x 14 inches 36 x 20 x 36 centimeters |
Workspace Volume: | 0.89 ft³ 25 L |
Temperature Range: | -40°C to 130°C (-40°F to 266°F) |
Temperature Control Tolerance: | ±1.1°C (2°F) |
Approx. Shipping Weight: | 250 lbs 113 kg |
Move-In Dimensions—WxDxH: | 18 x 23 x 30 inches 46 x 58 x 76 centimeters |
Humidity Range: | 20% to 95% |
Humidity Control Tolerance: | =+/-2.5% RH |
Heating Performance in Minutes 25º to 110ºC: | 16 |
Heating Performance in Minutes 25º to 65ºC: | 6 |
Electrical Requirements in Full Load Amps 220 / 1 / 50: | 9.5 |
Cooling Performance in Minutes 25º to -10ºC: | 11 |
Cooling Performance in Minutes 25º to -20ºC: | 22 |
Compressor Size | 1/3 HP |
Notes
- Humidity range is limited by a 7°C (45°F) minimum dewpoint and a maximum dry bulb of 85°C (185°F). SM-1.0 Benchtop Environmental Chamber and SM-3.5-8200 Benchtop Environmental Chamber are limited by 15°C (59°F) minimum dewpoint.
- Humidity control tolerance is measured at a dry bulb temperature above 20°C (68°F).
- Cooling and heating performance times are an average.
- S/SM-1.0 Benchtop Environmental Chamber with 115 / 1 / 60 electrical service has a 20 Amp minimum service/20 Amp plug on a 6 ft cord.
- Temperature Control Tolerance is ±1.1°C (2°F)
- The addition of accessories may impact performance or increase move-in dimensions.
- Performance is based upon 60 Hz and 23.9°C (75°F) ambient air. Benchtop Environmental Chambers are designed for normal, non-hazardous laboratory operating conditions. If hazardous materials are involved, please consult your local sales person.
- Custom options are available.
- Specifications are subject to change without notice.