SE-1000-6 Environmental Chamber

Workspace Dimensions—WxDxH: 40 x 39.25 x 38.25 inches
102 x 100 x 97 centimeters
Workspace Volume: 34.8 ft³
986 L
Temperature Uniformity: ±0.5ºC (±0.9ºF)
Temperature Range: -40ºC to 180ºC
(-40ºF to 356ºF)
Temperature Control Tolerance: 0.3°C (0.5°F)
Move-In Dimensions—WxDxH: 49 x 83 x 87 inches
124 x 211 x 221 centimeters
Inlet Water Gal / Liters per Minute 24ºC / 75ºF: Air-Cooled
Inlet Water Gal / Liters per Minute 18ºC / 65ºF: Air-Cooled
Inlet Water Gal / Liters per Minute 13ºC / 55ºF: Air-Cooled
Inlet Water Gal / Liters per Minute 29ºC / 85ºF: Air-Cooled
Humidity Range: 10 to 98% RH
Humidity Control: +/- 2.5% RH
Heating Performance in Minutes -35º to 71ºC: 15
Heating Performance in Minutes -35º to 180ºC: 36
Heating Performance in ºC/Minute -35º to 71ºC: 7ºC
Heating Performance in ºC/Minute -35º to 180ºC: 5.9ºC
Heating Performance in ºC/Minute-20º to 85ºC: 7ºC
Electrical Requirements in Full Load Amps 460 / 3 / 60: 30
Electrical Requirements in Full Load Amps 400 / 3 / 50: 31
Electrical Requirements in Full Load Amps 230 / 3 / 60: 59
Electrical Requirements in Full Load Amps 208 / 3 / 60: 61
Cooling Performance in Minutes 85º to -20ºC: 22
Cooling Performance in Minutes 71º to -35ºC: 33
Cooling Performance in Minutes 180º to -35ºC:52
Cooling Performance in ºC/Minute 85º to -20ºC: 4.7ºC
Cooling Performance in ºC/Minute 71º to -35ºC: 3.2ºC
Cooling Performance in ºC/Minute 180º to -35ºC: 4.1ºC
Airflow 1,000 CFM

Notes

  • The window sizes for all SE Environmental Chambers are 15 x 19 inches / 38 x 48cm. Custom window sizes are available.
  • Temperature Control: ±0.3ºC (±0.5ºF). Standard deviation from mean, measured at -25°C (-13°F) or 100°C (212°F).
  • Air temperature is measured with an empty climatic chamber
  • All SE-1000 Environmental Chambers are tested with 50 lbs/23 kg aluminum sheets.
  • Climatic Chamber performance is based upon laboratory ambient conditions of 23.9°C, and may vary slightly.
  • IEC specification based on the time the test chamber takes to pass through the middle 80% of the full temperature range when conducting a transition over this range.
  • Noise Level: A weighted sound pressure level measured at a distance of 1.0 meter (39.4 inches) from the equipment surface and a height of 1.6 meters (63 inches) from the floor in free-field conditions, using a calibrated instrument.
  • Air-cooled indicates that an onboard condenser is standard and a water-cooled condenser is optional. Environmental Chambers with listed water requirements can be built with remote air-cooled condensers.
  • Thermotron environmental test equipment is not designed to process hazardous materials. Consult an application engineer if hazardous materials are involved.

Additional accessories may impact performance of the test chamber. SE Environmental Chamber specifications are subject to change without notice.