Home  / 	 Environmental Test Equipment  / 	 Temperature & Humidity  / 	 S/SM Chambers  / 	 S/SM-Series Detail  / 	 SM-16-2-3-8200 Humidity Chamber | Performance Series
SM-16-2-3-8200 Humidity Chamber | Performance Series
| Workspace Dimensions—WxDxH: | 30 x 30 x 30 inches 76 x 76 x 76 centimeters | 
| Workspace Volume: | 16 ft³ 453 L | 
| Temperature Uniformity: | All Models 0.7°C (1.1°F) | 
| Temperature Range: | -70°C to +180°C (-94°F to +356°F) | 
| Temperature Control Tolerance: | All Models 0.3°C (0.5°F) | 
| Move-In Dimensions—WxDxH: | 45 x 59 x 81 inches 114 x 150 x 206 centimeters | 
| Humidity Range: | -68°C to 180°C (-90°F to 356°F) | 
| Heating Performance in Minutes -65º to 71ºC: | 10 | 
| Heating Performance in Minutes -65º to 180ºC: | 29 | 
| Heating Performance in Minutes -40º to 85ºC: | 9 | 
| Electrical Requirements in Full Load Amps 230 / 3 / 60: | 53 | 
| Electrical Requirements in Full Load Amps 230 / 1 / 60: | N/A | 
| Cooling Performance in Minutes 85º to -40ºC: | 17 | 
| Cooling Performance in Minutes 71º to -65ºC: | 36 | 
| Cooling Performance in Minutes 180º to -65ºC: | 53 | 
| Compressor Size | 2-3 Hp | 
Notes
- Humidity range in SM Series Humidity Chambers is limited by a +7°C (+45°F) minimum dew point temperature and a maximum dry bulb temperature of+88°C (+190°F). Relative humidity indication at or near the physical limits may be affected by sensor accuracy and control tolerance.
- Humidity Control Tolerance is measured at a dry bulb temperature above +20°C (+68 °F).
- Noise Level: A-weighted sound pressure level measured at a distance of 1.0 meter (39.4 inches) from the surface of the equipment at a height of 1.6 meters (63 inches) from the floor in free-field conditions, using a calibrated instrument.
- Performance is based upon 60Hz and 23.9°C (75°F) ambient air, and may vary slightly at other ambient temperatures. S/SM-Series Environmental Chambers are designed for use under normal laboratory operating conditions. For other applications, please consult Thermotron.
- Test Chamber specifications are subject to change.