Home  / 	 Environmental Test Equipment  / 	 Temperature & Humidity  / 	 S/SM Chambers  / 	 S/SM-Series Detail  / 	 SM-16-2-2-8200 Humidity Chamber
SM-16-2-2-8200 Humidity Chamber
| Workspace Dimensions—WxDxH: |  
     30 x 30 x 30 inches  76 x 76 x 76 centimeters  | 
| Workspace Volume: |  16 ft³  453 L  | 
| Temperature Uniformity: | All Models 0.7°C (1.1°F) | 
| Temperature Range: |  -68°C to +180°C  (-90°F to +356°F)  | 
| Temperature Control Tolerance: | All Models 0.3°C (0.5°F) | 
| Approx. Shipping Weight: |  1,395 lbs  633 kg  | 
| Move-In Dimensions—WxDxH: |  45 x 59 x 81 inches  114 x 150 x 206 centimeters  | 
| Humidity Range: | All Models 10% to 98% RH | 
| Humidity Control: | All Models 2.5% RH | 
| Heating Performance in Minutes -65º to 71ºC: | 22 | 
| Heating Performance in Minutes -65º to 180ºC: | 54 | 
| Heating Performance in Minutes -40º to 85ºC: | 21 | 
| Electrical Requirements in Full Load Amps 230 / 3 / 60: | 33 | 
| Electrical Requirements in Full Load Amps 230 / 1 / 60: | 46 | 
| Cooling Performance in Minutes 85º to -40ºC: | 31 | 
| Cooling Performance in Minutes 71º to -65ºC: | 54 | 
| Cooling Performance in Minutes 180º to -65ºC: | 80 | 
| Compressor Size | 2-2 Hp | 
Notes
- Humidity range in SM Series Humidity Chambers is limited by a +7°C (+45°F) minimum dew point temperature and a maximum dry bulb temperature of+88°C (+190°F). Relative humidity indication at or near the physical limits may be affected by sensor accuracy and control tolerance.
 - Humidity Control Tolerance is measured at a dry bulb temperature above +20°C (+68 °F).
 - Noise Level: A-weighted sound pressure level measured at a distance of 1.0 meter (39.4 inches) from the surface of the equipment at a height of 1.6 meters (63 inches) from the floor in free-field conditions, using a calibrated instrument.
 - Performance is based upon 60Hz and 23.9°C (75°F) ambient air, and may vary slightly at other ambient temperatures. S/SM-Series Environmental Chambers are designed for use under normal laboratory operating conditions. For other applications, please consult Thermotron.
 - Test Chamber specifications are subject to change.